The Clean Room Panel, which is fitted with a diaphragm sensor for differential-pressure measurement, is a top-quality device. Validated according to GMP directives, and with its well thought-out design, the CRP5 is optimized for applications in laboratories and clean rooms.
Clean Room Panel CRP5
Description
The Clean Room Panel, which is fitted with a diaphragm sensor for differential-pressure measurement, is a top-quality device. Validated according to GMP directives, and with its well thought-out design, the CRP5 is optimized for applications in laboratories and clean rooms. Optical buttons, and the magnet-mounted, removable HC2-CRP humidity and temperature probe permit efficient and effective cleaning. Up to 6 measurement values and messages can be displayed via its graphical color display. The colors of the display can be adjusted individually. Alarms, such as faults, exceeded limit values, or warnings are highlighted on the CRP5 display, and can also be forwarded digitally via MODBUS, Ethernet, or relays. Thanks to its analog and digital communication capabilities, the CRP5 is very easy to integrate in any monitoring system. Its comprehensive functional capabilities allow the CRP5 to be configured perfectly for each application. The CRP5 is distinctive in the marketplace through its tough glass front plate and a removable humidity sensor.
Applications
The CRP5 clean-room panel is the obvious choice wherever a high degree of cleanliness is vital, in addition to precise measurement of differential pressure, humidity, and temperature. The CRP5 is used particularly in clean-rooms in hospitals, in the pharmaceutical, electronics, and foodstuffs industries, and wherever small variances in pressure are monitored. The very comprehensive functional capabilities of the CRP5 provide solutions for many measurement and control tasks.
Features
- Measures differential pressure, humidity and temperature
- Designed specifically for clean rooms
- Removable humidity and temperature probes for simpler adjustment and cleaning
- FDA- and GAMP-compatible
- Digital communication via Ethernet RJ45 and/or MODBUS TCP / RTU
- Analog input and output signals freely configurable
- Glass front panel highly resistant to chemicals
- Optical operating elements for operation with protective gloves
Specifications
Probes | HC2-CRP; alternatively: any HygroClip2 probe |
Parameters | Differential pressure, temperature, relative humidity, digital and analog inputs |
Dimensions | 180 x 300 x 72 mm |
Weight | 1,700 g (with humidity sensor) 1,550 g (without humidity sensor) |
Range of application | -5…60 °C (23…140 °F) / 0…100 %RH |
Accuracy at 23 ±5 °C | ±1.0 % full-scale (Pa) |
Measurement principle | Diaphragm sensor |
Measurement ranges | ±50 Pa / ±100 Pa / ±250 Pa / ±500 Pa & -100…200 °C (probe-dependent) / 0…100 %RH |
Pressure resistance | 0.7 bar (70,000 Pa) |
Pressure connections | Front: Hose connectors Ø 6 mm x 10 mm Back: Hose connectors Ø 4 mm x 10 mm |
Medium | Air and non-aggressive gases |
Accuracy analog output | ±5 mV (voltage output) ±20 μA (current output) |
Switch output | 6 solid-state relays |
Switching capacity | <50 VDC at <2 A when observing polarity <35 VAC at <1 A, irrespective of polarity |
Zero drift | Compensated (manual or automatic zero adjustment) |
Power supply / Power consumption | 20…48 VDC / 18…35 VAC <500 mA |
Electrical connections | Screw terminals on back |
Measurement interval | 1 s |
Display type | Color TFT display |
Menu Navigation | 4 optical buttons |
Firmware update | Via HW4 software |
Analog outputs | 4, freely configurable |
Analog output signal (standard) | 0 / 4…20 mA or 0…1 / 5 / 10 V |
Permissible load | >10 kΩ (voltage output) <500 Ω (current output) |
Communication interfaces | Ethernet (Modbus TCP, HW4) / RS-485 (Modbus RTU, HW4) |
Service Interface | UART |
Housing | Front: Glass Back: Stainless Steel |
IP protection rating | Front: IP65 (also with sensors removed) Back: IP20 |
Fire protection class | Complies UL94-HB |
Mounting position | Wall Mounting |
FDA / GAMP | FDA 21 CFR part 11 / GAMP5 |
CE / EMC conform | EMC Directive 2004/108/EC |
-5…60 °C (23…140 °F) / 0…100 %RH |