Rotronic HC2-IMxxx-M

Price on request Request a quote

This metal industrial sensor from Rotronic is made for high temperatures, demanding industrial environments and applications where hygiene plays an important role. It measures humidity and temperature, and calculates dew point.

Description

This metal industrial sensor from Rotronic is made for high temperatures, demanding industrial environments and applications where hygiene plays an important role. It measures humidity and temperature, and calculates dew point.

Applications
Food industry, pharmaceutical industry, drying processes, industrial production processes

Features.

  • Application range: -100…200 °C, (screw-in sensor; -50…200 °C) / 0…100 %RH
  • Digital interface (UART) and scalable analogue outputs, 0…1 V
  • Standard scalable outputs: 0…1 V = -40…60 °C / 0…100 %RH
  • Adjustable to 23 °C and 10, 35, 80 %RH

Specifications

Probe typeIndustrial metal probe
DimensionsØ15 x 130 mm / Ø15 x 280 mm / Ø15 x 430 mm / Ø15 x 580 mm
Weight260 g / 400 g / 540 g / 680 g
Operating range-100…200 °C / 0…100 %rh
Accuracy at 23 ±5 °C±0.8 %rh / ±0.1 K
Accuracy w. adjustment profile “standard”bij 23 °C / 10, 35, 80 %Rh ± 0,8 %Rh / ± 0,1 K
Power supply / Power consumption3.3 V ± 0.1V / ~4.5 mA
Long-term stability<1 %rh / year
Humidity sensorHygromer IN-1
Temperature sensorPT100 1/3 DIN Class B
Probe protectionFilter carrier, brass nickel plated
Filter cartridgedepending on the application, order separately
Response time t 63<15 sec.
Allowed wind speed3 m/s
Psychrometric calculationsDew or frost point
Reproducibility Airchip<0.02 %rh / 0.01 K
Exploitation softwareROTRONIC HW4
Probe adjustment by software1 Point & multipoint %rh & °C
Firmware updateWith internet and HW4 software
User informationfrom Software HW4 (devicemanager)
Data processing via HW4with interface cables
Device Protection by passwordYes
Type of output signals2 x 0…1 VDC
Analogue outputs scaleable by the userYes
Analog output signal (standard)0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital output signal (standard)UART
Service InterfaceUART
Reverse voltage protectionMechanically in the connector
Max. cable lengthUp to 5 m with passive cable. Up to 100 m with active cable.
Probe materialStainless steel 1.4305 (V2A)
Probe cable materialFEP, black
Enclosure materialStainless steel 1.4305 (V2A)
FDA/GAMP compatibilityFDA 21 CFR Part 11 and GAMP compatible

Documentation

Brochure Hygroclip 2 Rotronic HC2-IMxxx-M




    Cart0
    There are no products in the cart!
    Continue shopping