Rotronic HygroLog HL-NT3

Price on request Request a quote

HygroLog NT data loggers have a wide range of applications, including the pharmaceutical, food and chemical industries. The data loggers are very accurate and easy to use and maintain. The HL-NT3 has two additional external sensing inputs. Additional sensors can be connected to the logger via a docking station.

Description

HygroLog NT data loggers have a wide range of applications, including the pharmaceutical, food and chemical industries. The data loggers are very accurate and easy to use and maintain. The HL-NT3 has two additional external sensing inputs. Additional sensors can be connected to the logger via a docking station.

The HygroLog NT uses Hygroclip digital humidity technology and this ensures uninterrupted measurement. Depending on the sensor used, the measuring range is between -50°C..200°C/ 0..100%RV with an accuracy of ±1.0%RV/ ±0.2°C. With the software HW4, the logger is FDA CFR 21.11 and GAMP compliant and ideally applicable in an environment where quality control is important.

Features.

  • Fixed or replaceable sensors
  • Integrated display (optional)
  • Alarm function
  • Easy to use keypad or software
  • Docking stations for extended options
  • Te gebruiken in een netwerk

Specifications

Probe typeStandard probe
Dimensions128 x 109 x 42 mm
Operating range-10…60 °C / 0…100 %rh
Accuracy at 23 ±5 °C±0.8 %rh / ±0.1 K
Measuring range with integrated probe-100…200 °C / 0…100 %rh
Power supply / Power consumptionvia optional docking station
Long-term stability
Humidity sensorHygromer IN-1
Temperature sensorPT100 1/3 DIN Class B
Probe protectionABS
Response time t 63Typical 4 sec.
Psychrometric calculationsAll psychrometric values
Number of probe inputs3
Input for temperature sensorOptional adapter cable
Calibration / adjustmentwith HW4 / docking station
Measurement intervalbetween 5 s and 24 h
Engineering units displayedValue display, status display, operation, adjustment and view the log data.
Backlightwith Backlight
Keypad (with display only)Yes
AlertingLED status indicator and audible alarm (beeper)
Acoustical alarm (Beeper)Yes
Real time clock w.backup batteryYes
Memory sizeMemory up to 47,000 records per MB on memory cards
Exploitation softwareROTRONIC HW4 / card reader, PDA
Menu guidanceYes
Trend indicatorYes, for models with display
Probe adjustment by softwareYes, with service cable
Probe adjustment by keypaddisplay / keyboard
Event loggingvia software HW4
Status indicator LEDYes
User informationfrom Software HW4 (devicemanager)
Data processing via HW4Yes
Device Protection by passwordYes
External dataloggingvia software HW4 / PC
Country settingsvia software HW4
Choice of Battery / rechargeable batteryOperation with a rechargeable battery or battery
Communication interfacesvia docking station
Service InterfaceUSB
Probe materialPolycarbonate
Enclosure materialABS
Anti theft deviceOptional with Docking Station
FDA/GAMP compatibilityFDA 21 CFR Part 11 and GAMP compatible