Multi Channel Voltage Recorder

Multichannel Recorder MCR-4V is a 4-channel battery operated voltage data logger. Up to 4 units can be coupled together, making it possible to simultaneously measure and record up to 16 channels.

Quick and precise measurements are made possible with a shortest recording interval of 2 msec, and a minimum voltage resolution of 10 µV.
Besides using a USB connection to directly send recorded data to a PC, it is possible to use an SD card to transfer data from a logger to a PC. Data can also be viewed and checked on the handy LCD display.

Features

  • Capable of 4-channel measurements and recording with a single unit
  • Simultaneous measurement and recording for up to 4 units (16 channels) by coupling units.
  • Runs on 2 AA Alkaline Batteries
  • Minimum Resolution of 10µV and Input Range of ±±24V
  • Recording Interval as short as 2 msec
  • Use of SD Memory Card for Data Transfer to PC
  • Logging Capacity of 480,000 data readings
  • Scale and Unit Conversion for each channel
  • Electrical Isolation between channels
  • Use of Trend Graph for Checking Data in Real Time

Specificaties

Measurement Channels Voltage 4ch
Input Method Scanning Method, Differential Input, Each Channel Isolated
Input Impedance Approx. 1.1 MΩ
Input Frequency DC – 100 Hz
Measurement Range ±300 mV, ±1.5 V, ±6 V, ±24 V, Auto (*1)
Absolute Maximum Input Voltage: ±50 V
Accuracy When the 50-60 Hz filter is ON, varies with the Measurement Range as follows:
±300 mV : ±(0.3 % + 0.06 mV)
±1.5 V : ±(0.3 % + 0.3 mV)
±6 V : ±(0.3 % + 0.6 mV)
±24 V : ±(0.3 % + 2.4 mV)
Auto : According to the range in use
Measurement Resolution 50-60 Hz Filter ON : 0.01 mV
50-60 Hz Filter OFF : 0.1 mV
Preheat Function Allowable Voltage Range : 3 to 24 V (external power supply)
Allowable Maximum Current : 1.0 A
Recording Method Instantaneous, Average, or Ave. Fine
Recording Interval 2, 5, 10, 20, 50, 100, 200, 500 ms.
1, 2, 5, 10, 15, 20, 30 sec.
1, 2, 5, 10, 15, 20, 30, 60 min.
(The minimum interval will depend on the number of channels, measurement range, and 50-60 Hz filter setting.)
Logging Capacity (*2) When recording 1 channel : up to 480,000 readings/ch
When recording 2 channels : up to 240,000 readings/ch
When recording 3 channels : up to 160,000 readings/ch
When recording 4 channels : up to 120,000 readings/ch
Recording Mode Endless (Overwrite the most recent data when capacity is full)
One Time (Stop recording when capacity is full)
Recording Start Method Immediate Start or Programmed Start (by individual logger or by group)
Recording Stop Trigger ON or OFF
Group Recording Up to 4 units (16 channels) can be recorded simultaneously.
LCD Display Items Measurements, Trend Graph, Battery Level, etc.
Communication Interfaces USB Communication
Communication Time Download Times for Full Data via USB
– While recording (at recording interval of 2 ms.) : Approx. 3 minutes 30 seconds
– With recording stopped: Approx. 1 minute 30 seconds
– From a slave unit: Approx. 4 minutes 30 seconds
External Memory SD Memory Card, SDHC Memory Card (*3)
Power AA Alkaline Battery x 2 (AA Ni-MH batteries may also be used), USB Power (5 V 250 mA)
Battery Life (*4) Approx. 4.5 to 150 days (with AA alkaline batteries)
Input Terminal /
Preheat Terminal
Screwless Terminals
< Compatible Wires >
Single Wire : φ 0.32mm to φ 0.65mm (AWG 28 – 22)
Twisted Wire : 0.08mm2 to 0.32mm2 (AWG 28 – 22), φ 0.12 mm or more in diameter
Stripping Length : 9 to 10 mm
Isolation CH1, CH2, CH3, CH4, USB, and Preheat are isolated. (Battery terminals are not isolated from the CH1-CH4 input terminals.)
CH1-CH4 Maximum Applied Voltage : ± 50 V
Electrical Isolation Resistance : 50 MΩ or more (DC ± 250 V)
Dimensions H 120 mm x W 75 mm x D 32 mm
Weight Approx. 190 g (including batteries)
Operating Environment Temperature: 0 to 50 °C
Humidity: 90 %RH or less (no condensation)
Accessories AA Alkaline Battery x 2, USB Communication Cable (US-15C), Software (CD-ROM), Card Slot Cover, User’s Manual Set (Warranty Included)