Rotronic HC2-HPxx

Price on request Request a quote

The HC2-HP probes are hand-held ø10mm probes. There are 2 different lengths available, 280mm and 500mm. The filter at the end is replaceable and thus ensures a longer life for the sensor.

Description

The HC2-HP probes are hand-held ø10mm probes. There are 2 different lengths available, 280mm and 500mm. The filter at the end is replaceable and thus ensures a longer life for the sensor.

Applications
10 mm insertion probe: measurements in dusty bulk materials such as flour, sugar, etc.

Features.

  • Application range: -40…85 °C / 0…100 %RH
  • Digital interface (UART) and scalable analogue outputs, 0…1 V
  • Standard scalable outputs: 0…1 V = -40…60 °C / 0…100 %RH
  • Adjustable to 23 °C and 10, 35, 80 %RH

Specifications

Probe typeInsertion probe
Dimensions280 mm / 500 mm
Weight200 g
For use withHygroLab C1, HP23-AW-A
Operating range-40…85 °C / 0…100 %rh
Range of application-40…85 °C (-40…185 °F) / 0…100 %RH
Accuracy at 23 ±5 °C± 0.008 aw / 0.8 %rh / ±0.1 K
Accuracy w. adjustment profile “standard”bij 23 °C / 10, 35, 80 %Rh ± 0,8 %Rh / ± 0,1 K
Power supply3.2…5 VDC, calibrated at 3.3 VDC, current: ~4.5 mA
Long-term stability
Humidity sensorHYGROMER® WA-1 / Pt-100, 1/3 DIN Class B
Temperature dependence0.001 K/K, 0.0002 aw/K
Temperature sensorPT100 1/3 DIN Class B
Filter typeSintered steel with 5 μm pore size
Response time t 63
Allowed wind speed40 m/s, with filter
Psychrometric calculationsDew or frost point
Reproducibility Airchip
Start-up time1,5 s
Measurement interval1 s
Exploitation softwareROTRONIC HW4
Probe adjustment by software1 Point & multipoint %rh & °C
Firmware updateWith internet and HW4 software
User informationfrom Software HW4 (devicemanager)
Data processing via HW4with interface cables
Device Protection by passwordYes
Type of output signals2 x 0…1 VDC
Analogue outputs scaleable by the userYes
Analog output signal (standard)0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital output signal (standard)UART
Service InterfaceUART
Reverse voltage protectionMechanically in the connector
Max. cable lengthUp to 5 m with passive cable. Up to 100 m with active cable.
Probe materialStainless steel 1.4305 (V2A)
Enclosure materialPOM
MaterialStainless steel DIN 1.4305 (probe), POM (handle)
IP protection ratingIP50
FDA/GAMP compatibilityFDA 21 CFR Part 11 and GAMP compatible
FDA / GAMPFDA 21 CFR part 11 / GAMP5
CE / EMC conform2007 / 108 / EG

Documentation

Brochure Hygroclip 2 Rotronic HC2-HPxx




    Cart0
    There are no products in the cart!
    Continue shopping
    0